Injection speed |
The 10ml syringe speed setting range is within (0.1 ∼300.0) ml/h, and the minimum step is 0.1ml/h. The speed setting range of the 20ml syringe is within (0.1 ∼600.0) ml/h, and the minimum step is 0.1ml/h. The 30ml syringe speed setting range is within (0.1 ∼900.0)ml/h, and the minimum step is 0.1ml/h. The 50ml (60ml) syringe speed setting range is within (0.1 ∼1500.0) ml/h, and the minimum step is 0.1ml/h |
|
Fast bolus function |
1)The 10ml syringe rapid injection speed setting range is (0.1 ∼240.0) ml/h, the injection volume setting range is (0.1 ∼5.0) ml, and the minimum step is 0.1ml/h. 2)The rapid injection speed setting range of 20ml syringe is within (0.1∼480.0)ml/h, the injection volume setting range is (0.1 ∼10.0)ml, and the minimum step is 0.1ml/h. 3)The 30ml syringe rapid injection speed setting range is (0.1 ∼720.0) ml/h, the injection volume setting range is (0.1∼10.0) ml, and the minimum step is 0.1ml/h. 4)The 50ml (60ml) syringe rapid injection speed setting range is (0.1 ∼1200.0) ml/h, the injection volume setting range is (0.1 ∼10.0) ml, and the minimum step is 0.1ml/h. |
|
Infusion accuracy |
Injection error≤±3% |
|
Limit amount setting range |
Can be in the range of (0.1∼9999.9)ml, the smallest step is 0.1ml |
|
Cumulative infusion volume display |
The syringe pump can display the cumulative amount during the injection process, the display range is (0.1∼9999.9) ml range, the minimum step is 0.1ml |
|
KVO speed |
KVO speed (0.1 ∼5.0) ml/h adjustable, the minimum step is 0.1ml |
|
Anti-bolus |
When the pipeline is blocked, the pipeline pressure can be automatically reduced. |
|
Syringe recognition |
The syringe pump can automatically recognize four types of syringes: 10ml, 20ml, 30ml, 50ml (60ml) |
|
Alarm type |
Injection block alarm/Syringe falling off alarm/Injection near alarm/Injection complete alarm/Missing operation alarm/Injector not properly installed alarm/Low battery power alarm/Battery empty alarm/Internet power interruption prompt/Injection speed setting error prompt/Injection start prompt . |
|
power supply |
Three power supply modes are optional 1)AC100 ∼240V 50/60HZ 2) Built-in lithium battery 11.1V, 1800mAh, after the battery is fully charged, it can work continuously for more than 2.5 hours (5ml/h) 3) Support car, can use 12V power supply in car |
|
Medicine library function |
With drug library function, there are more than 1,000 drug libraries in more than 10 categories |
|
Product Size |
304mm*174mm*260mm |